Ceraphant T PTC31, PTP31, PTP35
Trouble-shooting
Hauser
27
7.2
Repair
A repair is not planned.
7.3
Disposal
When disposing, ensure that the materials of the device components are separated and processed
accordingly.
7.4
Change status (release)
The release number on the nameplate and in the Operating Instructions indicates the change
status of the device: X.YY. (example 1.02. ).
W-Code
Explanation
Remedy
W107
Simulation active
Switch off the output simulation for output 1
and output 2
W202
Pressure outside the sensor range
Operate the device in the specified pressure
range
W209
Device starts
W210
Configuration modified (warning code will be
displayed for 15 s approx.)
W212
Sensor signal outside the permitted range
Operate the device in the specified pressure
range
W250
Number of switch cycles exceeded
Replace the device
W270
Short-circuit and overload at output 1
Check output wiring. Extend the load
resistance at output 1
W280
Short-circuit and overload at output 2
Check output wiring. Extend the load
resistance at output 2
X
Change in the main version.
Compatibility no longer provided. Device and Operating Instructions change.
YY
Compatibility provided. Operating Instructions change.