Cerabar M
62
Hauser
Note!
• The influence of temperature on the zero point can be corrected through position adjustment.
• The temperature influence can be minimized by using a filling oil with a smaller expansion coefficient, a
shorter capillary, a diaphragm seal with a larger diaphragm diameter or by using a smaller capillary internal
diameter.
P01-PMP46xxx-05-xx-xx-xx-002
Diagram T
K
Ambient as a function of the capillary length for PMP46
Characteristic
type
Version
1)
1)
Version for feature 70 "Process connection" in the order code
Diaphragm seal
➀
DU
Tri-Clamp, ISO 2852 DN 76.1 (3"), AISI 316L
➁
SL
Pipe diaphragm seal Tri-Clamp, ISO 2852 DN 51 (2"), AISI 316L
➂
SG
Pipe diaphragm seal Tri-Clamp, ISO 2852 DN 38 (1 1/2"), AISI 316L
➃
AL
DIN 11851 DN 50 PN 25, AISI 316L
PL
Pipe diaphragm seal DIN 11581 DN 50 PN 25, AISI 316L
➄
LL
Varivent type N for pipes DN 40 – DN 162, PN 40, AISI 316L
KL
DRD DN50 (65 mm) PN 25, AISI 316L
10
11
9
6
0
20
40
60
7
0
8
0
90
100
10
30
50
0
1
2
3
4
5
6
7
8
9
10
7
8
12
[mbar
/
10 K]
T Ambient
K
[m]
Capillary length
Characteristic type
Version
1)
Diaphragm seal
➅
DL
Tri-Clamp, ISO 2852 DN 51 (2"), DIN 32676 DN 50, AISI 316L
➆
2)
SB
Pipe diaphragm seal Tri-Clamp, ISO 2852 DN 25 (1"), AISI 316L
LB
Varivent type F for pipes DN 25 – DN 32 PN40, AISI 316L
Summary of Contents for Cerabar M PMC41
Page 83: ...Cerabar M Endress Hauser 83...