18
Hauser
General warnings
• Compliance with national regulations relating to the installation, connection to the electricity supply,
commissioning and maintenance of devices in potentially explosive atmospheres is mandatory, if such
regulations exist.
• Installation, connection to the electricity supply, commissioning and maintenance of the devices must be
carried out by qualified specialists trained to work on Ex-rated devices.
• Compliance with all of the technical data of the device (see nameplate) is mandatory.
• Open the device only when it is de-energized (and after a delay of at least 10 minutes following shutdown
of the power supply).
• All seals on the transmitter housing must be checked at suitable time intervals to ensure that they are leak-
tight and they must be replaced, if necessary. When replacing the seals, use only original seals from
Hauser.
• It is not permissible to connect the service adapter whilst the atmosphere is considered to be explosive.
• Use of the devices is restricted to mediums against which the process-wetted materials are adequately
resistant.
Installation instructions
• The measuring device must only be used in the permitted temperature class.
The values of the individual temperature classes can be found in the temperature tables on Page 21.
• Turning the transmitter housing: the housing of the Ex-rated transmitter must only be turned in 90° steps
when the transmitter is de-energized (and after a delay of at least 10 minutes following shutdown of the
power supply).
• Turning the local display: the screw cap has to be removed before the local display can be turned, and this
must be done with the device de-energized (and after a delay of at least 10 minutes following shutdown of
the power supply)
• The cable entries and openings not used must be sealed tight with suitable components.
EC type-examination
certificate,
directive 94/9/EC
General
The system meets the fundamental health and safety requirements for the design and construction of devices
and protective systems intended for use in potentially explosive atmospheres in accordance with directive
94/9/EC as per EN 50021.
Inspection body
The transmitter housing has been tested for protection from vapors by
→
TÜV Nord Anlagentechnik GmbH
Identification
• The identification of the transmitter contains the following specifications:
–
Ð0
II3G EEx nRC IIC T6-T1 X (compact version)
or
–
Ð0
II3G EEx nRC IIC T6 X (remote version)
• The identification of the sensor contains the following specifications:
–
Ð0
II3G EEx nA II T6-T1 X
"
Caution!
The installation instructions for safe application of the system must be observed.