ZEEnit 700 Q
Safety instructions
9
2.4
Safety instructions, transport and installation
Observe the following notes:
The ZEEnit 700 Q is always installed by the service department of Analytik Jena or
by specialist personnel that was authorized and trained by Analytik Jena. Inde-
pendent assembly and installation are not permitted. Incorrect installation can cre-
ate serious hazards.
The ZEEnit 700 Q weighs 225 kg. Use a lift truck to transport the device.
Four people are required to move the device in the laboratory by holding the device
on four firmly screwed-in carrying handles.
Perform a professional and documented decontamination of the device before
returning it to Analytik Jena. The decontamination protocol is provided by the ser-
vice when the return is registered. Analytik Jena must refuse acceptance of contam-
inated devices. The sender may be liable for any damage caused by inadequate de-
contamination of the device.
The ZEEnit 700 Q must not be operated in an explosive environment.
Smoking and open flames in the operating room of the AAS are prohibited!
The operator is responsible for establishing a control method to ensure that the
N
2
O and acetylene connectors are leak-tight.
2.5
Safety instructions for operation
Prior to starting up the device, the operator of the ZEEnit 700 Q must verify the
proper condition of the device including the condition of its safety equipment. This
applies in particular after each modification or extension of the device or its repair.
Modifications and extensions to the device may only be carried out after consulta-
tion with Analytik Jena. Unauthorized changes can limit the operating safety and
lead to restrictions in warranty and access to customer service.
The device must only be operated if all protective equipment (e.g. covers and
doors) are in place, properly installed and fully operational. The sound condition of
the protection and safety equipment must be checked regularly. Any defects must
be corrected as soon as they occur. Protective and safety equipment must never be
removed, modified or switched off during operation.
During operation, the operator must have free access to all connections and the
power switch on the right side of the device as well as to the power strip.
The ventilation equipment on the device must be in good working condition. Cov-
ered ventilation grilles or slits etc. may cause the device to break down or damage
the device. When placing the device and system components observe a minimum
clearance of 150 mm to walls and neighboring installations.
Prevent any liquids from entering the inside of the instrument. The liquids might
get into contact with electronic components and cause a short circuit.
Protection against explo-
sion and fire