SPECORD PLUS
Specifications
43
10.2 General technical data of the SPECORD PLUS
Mass
SPECORD 50 PLUS: 21.0 kg
SPECORD 200 PLUS: 21.9 kg
SPECORD 210 PLUS: 22.2 kg
SPECORD 250 PLUS: 22.8 kg
Floor space (width x depth)
approx 900 x 690 mm with PC
Dimensions (width x height x depth)
590 x 290 x 690 mm
Sample chamber dimensions (width x height
x depth)
364 x 200 x 185 mm
Mains voltage
85 – 264 V/AC
Frequency
50 – 60 Hz
Power consumption
200 VA
Mains fuses
2 x T 3.15 AH / 250 V, type 19181 by Wick-
mann
Data connections
2 x RS-232 for accessories
1 x USB port for connecting an external PC
1 x RS-232 for connecting an external PC
Ambient conditions
Temperature during operation
15 to 35 °C
Humidity during operation
90 % at 30 °C
Air pressure
0.7 to 1.06 bar
Temperature and humidity during storage
15 – 55 °C at 10 – 30 % humidity, use dessi-
cant
Energy consumption
Energy consumption in different operating modes (SPECORD 210 PLUS)
Operating state
Power
in [VA]
Power
in [W]
Energy consumption
per year in [kW]
3 h / 5-day week
Energy consump-
tion per year in
[kW]
8 h / 5-day week
Standby without
lamps
31.3
16.6
12.9
34.5
Standby, VIS lamp
only
69.1
40.4
31.5
84.0
Standby, VIS lamp
only
92.5
54
42.1
112.3
Standby, VIS and UV
lamp
128
77.8
60.7
161.8
Both lamps on, scan
only in UV
131.5
82.7
64.5
172.0
Both lamps on, scan
only in UV
130.2
79.6
62.1
165.6
The power values for the SPECORD 50 PLUS are lower by approx. 10 VA / 7 W in all op-
erating states.
The power values for the SPECORD 200 PLUS are lower by approx. 10 VA / 7 W in all
operating states.