Care and maintenance
SPECORD PLUS
32
CAUTION
Risk of burns from hot lamps!
Allow the lamps sufficient cooling time before replacement.
CAUTION
Risk of eye damage from UV radiation
Do not look into the radiation of the lamp either directly or indirectly through a mirror!
The emitted UV radiation can result in eye damage.
NOTICE
Contamination can impair the lamp properties!
Do not touch the glass parts of the new lamps with your fingers. Exercise particular cau-
tion with the UV lamp and do not touch the exit window of the beam! If the glass has
been touched by fingers, wipe the lamps with a clean, lint-free cloth and pure alcohol.
The lamp chamber is located on the device rear behind the service flap.
Figure 20Lamps and the dessicant cartridge in the lamp chamber
1 Desiccant cartridge (optional)
2 Plug connector of the UV lamp
3 Knurled head screws for fastening the
UV lamp
4 UV lamp holder
5 Front VIS lamp cover
6 Knurled head screw for fastening the VIS
lamp cover
Lamp service life
The lamp service life is:
¡
Halogen lamp (VIS lamp): > 2000 h
¡
Deuterium lamp (UV lamp): > 2000 h
The lamp properties can be checked via the software. To do this, select menu item In-
strument | Test | Lamp check in ASpect UV.