Endress+Hauser 846-051-500 Manual Download Page 8

 
 

 

Manual Biometra TSC ThermoShaker, Vers. 01_2019 

 Installation 

 
6.1 

Unpack and Check 

 
Unpack and carefully examine the 

TSC ThermoShaker

. Report any damage to Analytik 

Jena. Save all packing material if damage is found. 
 
 

 

Attention

! Automatic balancing system in this product produces light metal-like 

noise  when  moving  the  unit  which  is  likely  to  be  heard  during  unpacking.  It  is 
normal occurrence and does not indicate to a fault! 

 
 
Do not attempt to operate this device if physical damage is present. 
 
If you would like to send the unit back to us, please read the return instructions (page 17). 

 
 
6.2 

Installation 

 
Place the apparatus upon even horizontal non-flammable surface away from any flammable 
material. 
 
To provide optimum ventilation ensure clearance around the apparatus: 20 cm on each side. 
Ensure additional clearance from the bottom side and do not place any object below the unit. 
 
Plug the external power supply unit (AC/DC adapter supplied with the shaker) into the 12 V 
socket at the rear side of the apparatus. 
 
 
 

6.3 

Installation of Block Modules 

 
 

 

Attention:

 Installation and change of block modules have to be performed 

 

 

only when the Power switch is turned off and the external power supply is 

 

 

disconnected from the mains outlet. 

 
Choose the block module. 
 
Connect the plug to the contact terminal according to Fig. 3/1 on the bottom side of the block 
module. Make sure that the connector is mounted tightly. 
 
 
 
 
 

Fig. 3: Connecting block modules 

 
 
Align the block module that the warning labels are facing the front of the unit. 
 
Secure with the four knurled screws. 
Installing the 96-well block module, drive one by one 4 screws for several turns and fix them 
with the hexagonal key (included with the block module. 
 
 
 

1

Summary of Contents for 846-051-500

Page 1: ...ing this apparatus Biometra GmbH Service Department Rudolf Wissell Str 30 Rudolf Wissell Str 14 16 D 37079 G ttingen D 37079 G ttingen Germany Germany Phone 49 5 51 50 68 6 0 Phone 49 5 51 50 88 1 10...

Page 2: ...LES 8 7 OPERATION OF TSC THERMOSHAKER 9 7 1 PARAMETER SETTING 10 7 1 1 Reaction Time TIME 10 7 1 2 Shaking Intensity RPM 10 7 1 3 Reaction Temperature T C 10 7 1 4 Program Execution 10 8 CALIBRATION 1...

Page 3: ...During drying out the intrinsic protection may be impaired The unit must be placed on a level non flammable surface away from flammable materials Use appropriate vessels tubes for temperature required...

Page 4: ...spilt inside the unit disconnect it from the power supply and have it checked by a competent person Disconnect the mains before removing the outer cover Note there are no user serviceable parts insid...

Page 5: ...51 600 690 is delivered with Fig 1 Delivered parts 1 x ThermoShaker apparatus Fig 1 1 1 x External power supply unit cable Fig 1 2 2 x Spare rubber belt Fig 1 3 4 x Rubber spacers for block screws 1 x...

Page 6: ...dule for TSC 20 x 0 2 ml 12 x 1 5 ml tubes 846 051 613 Interchangeable block module for TSC 20 x 0 5 ml 12 x 1 5 ml tubes 846 051 614 Interchangeable block module for 96 well microtiter plates or 96 x...

Page 7: ...r without temperature control The TSC ThermoShaker is applicable for DNA and RNA analysis extraction of lipids and other cell components DNA library creation biochemical studies of enzymatic reactions...

Page 8: ...ound the apparatus 20 cm on each side Ensure additional clearance from the bottom side and do not place any object below the unit Plug the external power supply unit AC DC adapter supplied with the sh...

Page 9: ...he mains outlet Remove the four knurled screws Fig 4 1 and disconnect the 10 way plug Fig 4 2 Fig 4 Changing block modules Select the new block module Install the block module according to 6 3 Attenti...

Page 10: ...lth risk when working with infected material To prevent such cases it is recommended to use tubes with cap lock of Safe Lock type For efficient mixing it is recommended to fill test tubes up to 75 of...

Page 11: ...r 10 RPM for the first 10 steps after these 100 RPM increments to reach 1 400 RPM 7 1 3 Reaction Temperature T C With the help of and buttons Fig 6 3 set the necessary temperature increment 0 1 C If t...

Page 12: ...e Timer Start Stop button once Fig 6 5 to stop the timer Press the Timer Start Stop button again to restart the timer The platform motion can be stopped at any time by pressing the Shaker Start Stop b...

Page 13: ...thermoblock special sensor socket Set the necessary temperature e g 40 C After the unit reaches the set temperature when the set and current temperature readings becomes the same it is necessary to l...

Page 14: ...ample the calibration coefficient will be 0 974 in range 0 936 up to 1 063 increment 0 001 This calibrating coefficient will correct temperature though all operation range After calibration press SHAK...

Page 15: ...Disconnect the unit from the power supply socket Before using any decontamination or cleaning method except that recommended check with our Service Department that the proposed method will not damage...

Page 16: ...in in the range from RT to 15 C below 25 C 1 8 C min from 25 C to 4 C in 12 min Time of block module heating from 25 C to 37 C 6 min 15 min average including temperature stabilisation Temperature cali...

Page 17: ...tput DC 12 V Dimensions W x D x H 205 x 230 x 130 mm Weight with power supply 4 1 kg at max The TS1 ThermoShaker is designed for operation in closed laboratory rooms cold rooms or incubators at ambien...

Page 18: ...tamination of the device Please clean all device components from biologically hazardous chemical or radioactive contamination see also section Cleaning Download the decontamination declaration as an e...

Page 19: ...e brought to the return systems and or separate systems available to end users according to yours country regulations when this product has reached the end of its lifetime For details please contact y...

Page 20: ...iometra TSC ThermoShaker Biometra TS1 ThermoShaker Best Nr Order No 846 051 600 846 051 690 846 051 500 846 051 590 den folgenden Europ ischen Richtlinien und angewandten harmonisierten Normen entspre...

Reviews: