Manual Biometra TSC ThermoShaker, Vers. 01_2019 

11 

At the end of the program (after the set time elapses) the platform motion stops and the timer 
shows the flashing reading STOP accompanied by the repetitive sound signal until the 
“Shaker”-Start/Stop button is pressed.  
 
If the working time is not set (or deleted) and the timer indicator in the upper line shows 
00:00, pressing the 

“Shaker”-Start/Stop button causes the apparatus to operate continuously 

until the “Shaker”-Start/Stop button is pressed again. 
 
If required, there is possibility to restart the timer when it is running. Press the 

“Timer”-

Start/Stop button once (Fig. 6/5) to stop the timer. Press the 

“Timer”-Start/Stop button again 

to restart the timer. 
 
The platform motion can be stopped at any time by pressing the 

“Shaker”-Start/Stop button. 

In this case the program and the platform motion stop, the timer is set back to zero and 
switches into the STOP mode. Press the 

“Shaker”-Start/Stop” button to repeat the operation 

with the same time and speed. 
 

Attention: 
 

At the end of the set time period the platform movement is stopped 
automatically, but the heating can be stopped only manually by reducing the 
temperature with the “▼“ T(ºC) key (Fig. 6/3 - lower button) till the OFF sign 
appears in the upper line of the display. 

 
 
 
At the end of operation switch off the unit by using the on-off switch at the rear side of the 
TS1 (Fig. 5). 
 
 

Summary of Contents for 846-051-500

Page 1: ...ing this apparatus Biometra GmbH Service Department Rudolf Wissell Str 30 Rudolf Wissell Str 14 16 D 37079 G ttingen D 37079 G ttingen Germany Germany Phone 49 5 51 50 68 6 0 Phone 49 5 51 50 88 1 10...

Page 2: ...LES 8 7 OPERATION OF TSC THERMOSHAKER 9 7 1 PARAMETER SETTING 10 7 1 1 Reaction Time TIME 10 7 1 2 Shaking Intensity RPM 10 7 1 3 Reaction Temperature T C 10 7 1 4 Program Execution 10 8 CALIBRATION 1...

Page 3: ...During drying out the intrinsic protection may be impaired The unit must be placed on a level non flammable surface away from flammable materials Use appropriate vessels tubes for temperature required...

Page 4: ...spilt inside the unit disconnect it from the power supply and have it checked by a competent person Disconnect the mains before removing the outer cover Note there are no user serviceable parts insid...

Page 5: ...51 600 690 is delivered with Fig 1 Delivered parts 1 x ThermoShaker apparatus Fig 1 1 1 x External power supply unit cable Fig 1 2 2 x Spare rubber belt Fig 1 3 4 x Rubber spacers for block screws 1 x...

Page 6: ...dule for TSC 20 x 0 2 ml 12 x 1 5 ml tubes 846 051 613 Interchangeable block module for TSC 20 x 0 5 ml 12 x 1 5 ml tubes 846 051 614 Interchangeable block module for 96 well microtiter plates or 96 x...

Page 7: ...r without temperature control The TSC ThermoShaker is applicable for DNA and RNA analysis extraction of lipids and other cell components DNA library creation biochemical studies of enzymatic reactions...

Page 8: ...ound the apparatus 20 cm on each side Ensure additional clearance from the bottom side and do not place any object below the unit Plug the external power supply unit AC DC adapter supplied with the sh...

Page 9: ...he mains outlet Remove the four knurled screws Fig 4 1 and disconnect the 10 way plug Fig 4 2 Fig 4 Changing block modules Select the new block module Install the block module according to 6 3 Attenti...

Page 10: ...lth risk when working with infected material To prevent such cases it is recommended to use tubes with cap lock of Safe Lock type For efficient mixing it is recommended to fill test tubes up to 75 of...

Page 11: ...r 10 RPM for the first 10 steps after these 100 RPM increments to reach 1 400 RPM 7 1 3 Reaction Temperature T C With the help of and buttons Fig 6 3 set the necessary temperature increment 0 1 C If t...

Page 12: ...e Timer Start Stop button once Fig 6 5 to stop the timer Press the Timer Start Stop button again to restart the timer The platform motion can be stopped at any time by pressing the Shaker Start Stop b...

Page 13: ...thermoblock special sensor socket Set the necessary temperature e g 40 C After the unit reaches the set temperature when the set and current temperature readings becomes the same it is necessary to l...

Page 14: ...ample the calibration coefficient will be 0 974 in range 0 936 up to 1 063 increment 0 001 This calibrating coefficient will correct temperature though all operation range After calibration press SHAK...

Page 15: ...Disconnect the unit from the power supply socket Before using any decontamination or cleaning method except that recommended check with our Service Department that the proposed method will not damage...

Page 16: ...in in the range from RT to 15 C below 25 C 1 8 C min from 25 C to 4 C in 12 min Time of block module heating from 25 C to 37 C 6 min 15 min average including temperature stabilisation Temperature cali...

Page 17: ...tput DC 12 V Dimensions W x D x H 205 x 230 x 130 mm Weight with power supply 4 1 kg at max The TS1 ThermoShaker is designed for operation in closed laboratory rooms cold rooms or incubators at ambien...

Page 18: ...tamination of the device Please clean all device components from biologically hazardous chemical or radioactive contamination see also section Cleaning Download the decontamination declaration as an e...

Page 19: ...e brought to the return systems and or separate systems available to end users according to yours country regulations when this product has reached the end of its lifetime For details please contact y...

Page 20: ...iometra TSC ThermoShaker Biometra TS1 ThermoShaker Best Nr Order No 846 051 600 846 051 690 846 051 500 846 051 590 den folgenden Europ ischen Richtlinien und angewandten harmonisierten Normen entspre...

Reviews: