Manual Biometra TSC ThermoShaker, Vers. 01_2019
13
Fig.9 Control panel in calibration and operation mode
To set the new temperature value, use the Temperature with multiplier - coefficient readings
(Fig.8/3).
With the help of Temp. “▲” and “▼“ keys (Fig.6/3) change the calibration coefficient
(Fig.9A/1) so, that the new temperature value (Fig.9A/2) corresponds to the independent
sensor temperature. In our example the calibration coefficient will be 0.974 (in range: 0.936
up to 1.063; increment 0.001).
This calibrating coefficient will correct temperature though all operation range.
After calibration press SHAKER-START/STOP key (Fig.3/4) once to save the changes and
exit calibration mode.
The display will show calibrated temperature as shown on fig.9B/3 and the unit will continue
thermal stabilization according to the previously set temperature.
calibrating
coefficient
sct. 40.0 40.0
0.974 39.0 40.0
00:00
40.0
1000
STOP
39.0
000
0
temp. with
coeff.
multiplier -
set temperature
calibrated temperature