2-Pack/3-Pack TDLAS Gas Analyzer
XA02751C
Hauser
7
Input power,
electronics enclosure
Analyzer B
120 VAC or /-10%, 50 to 60 Hz, 60W
max or 18 to 24VDC, 1.6 A maximum
Input power, Sample
Conditioning System
(SCS)
SCS Input Power - 120VAC or 240VAC, 200W
or 400W maximum
1
.
Analog
communication
Isolated Analog channels, 120 ohms at 24 VDC
maximum
Outputs: Qty 2 4-20 mA (measurement value)
Serial
communication
Channel 1 (H
S) - RS232C and Ethernet
Channel 2 and 3 (H
O and/or CO
) - RS232C or
Ethernet (TSP only)
Digital signals
Outputs: Qty 5 Hi/Lo alarm, general fault, vali
-
dation fail
2
, validation 1 active
, validation 2
active
Inputs: Qty 2 flow alarm
, validation request
Protocol
Modbus Gould RTU or Daniel RTU or ASCII
Diagnostic value
examples
Detector power (mirror health), spectrum ref
-
erence comparison and peak tracking (spec
-
trum quality), cell pressure and temperature
(overall system health)
LCD display
Concentration, cell pressure and temperature,
diagnostics
Physical Specifications
Electronics
enclosure type
Type 4X 304 or 316L stainless steel
Sample system
enclosure(s)
Type 4X 304 or 316L stainless steel
Analyzer dimensions
1285 mm H x 762 mm W x 394 mm D
(50.6 x 30 x 15.5 inches)
1. Application dependent
2. Configuration dependent