If the product is not operating normally, please refer to the following method for 

maintenance and adjustment.

There are no parts you can repair by yourself. If the machine still can not run normally after the above 

method, please contact the dealer or our company.

This device does not include accessories for repair usage; the repair should be carried out by an 

authorized person or authorized service center.

•  Keep the equipment in a dry storage condition.

•  Do not smear the equipment with pigments.

•  Calibration is recommended when using a new/other contra-angle handpiece.

•  Replace the battery if it seems to be running out of power sooner than it should.

•  Use soft cloth with ethanol to scrub the shell of the motor of the handpiece. Never use high 

concentration ethanol or corrosive liquid.

•  If you don’t usually use the product, please charge it often.

Can’t turning on

Troubles

Possible problems and treatment method

No display of charging

sign while charging

The auto-reverse does

not work

•  The batteries are probably dead, please charge.
•  Pressing  time of key is too short, please press the power key for 

over 1 second.

•  The power key does not work, please contact local dealer or 

manufacturer.

•  The power adapter is not plugged in, please reconnect the power 

adapter.

•  The charging base is not properly placed in the charging seat.
•  There is a foreign object on the bottom of the motor of the 

handpiece or on the cradle, please clean up.

Batteries are weakened (lower remaining battery capacity), please 

charge the batteries.

Troubleshooting Guide 

Maintenance

10.1 Lubrication of contra-angle handpiece

•  Lubrication should be limited to the contra-angle handpiece only. Apply lubricant spray after each 

use and/or before sterilization.

•  When you lubricate, please aim an injection nozzle at the rear part of the contra-angle and 

lubricate for 2-3 seconds. •  Do not lubricate the motor of the handpiece.

10.2 Cleaning motor of the handpiece

•  When the motor of the handpiece becomes dirty, wipe it off with a cotton cloth moistened with 

rubbing alcohol.

10.3 Cleaning charger

•  When the charger becomes dirty, wipe it off with a cotton cloth moistened with surgical spirit.

10.4 Sterilization

•  Sterilize the contra-angle handpiece only. For the sterilization method we recommend autoclaving.

•  Please use a brush (do not use metal brush) to brush away the dirt attached to the surface, and then 

wipe it with a clean soft cloth using surgical spirit.

•  Please place the contra angle handpiece into a sterilization bag and seal it.

•  The temperature should not exceed 135°C (eg 121°C for 20 minutes or 132°C for 15 minutes is ideal 

for autoclaving).

•  Keep the contra-angle handpiece in the autoclave pouch to keep it clean until you use it.

Enviromental Protection

11.1 Storage conditions

•  Storage temperature range : -10°C -  +50°C

•  Relative humidity range : less than 95%

•  Atmospheric pressure range : 50KPa - 106KPa

11.2 Environmental conditions

Do not expose to sun, rain, dust, corrosive gases or volatile chemicals keep the device in a ventilated 

place.

Disinfection

11

12

Summary of Contents for Endo Motor

Page 1: ...uct Manual Please read this manual carefully before use CORDLESS LED Endo Motor VASA DENTICITY PVT LTD Khasra No 714 Village P O Chattarpur Opp DLF Gate no 2 New Delhi 110074 India Phone No 91 7289999...

Page 2: ...se the machine safely to increase its reliability and performance operation 1 This machine is only for dental use Do not use it in other places 2 Standard configuration contra angle match for this mac...

Page 3: ...y which reduces costs for the machine Fully charged batteries generally discharge gradually over time even though the motor is not used It is recommended to place the handpiece in the charger when not...

Page 4: ...Setting TQ item Select key S speed Setting SP The selected Item will be a flashing display The speed and torque settings can be changed during the running of the machine The transfer sequence ditto To...

Page 5: ...aded will be reversed again Forward reciprocating motion Turn off automatic reverse no automatic reverse Tum off automatic forward no automatic forward Reverse the reciprocating motion the speed can b...

Page 6: ...to discharge the lamp holder 5 5 Operation Long press the Power Key the machine will be open and the screen will display Then click the motor switch files will begin to rotate and the light will be on...

Page 7: ...e please clean up Batteries are weakened lower remaining battery capacity please charge the batteries Troubleshooting Guide Maintenance 10 1 Lubrication of contra angle handpiece Lubrication should be...

Page 8: ...aintenance service free for six months If accessories such as Power Adapter Motor cover and Lamp Holder damage or run out you can purchase them from our company 11 3 Transportation Ordinary modes of t...

Reviews: