If the product is not operating normally, please refer to the following method for
maintenance and adjustment.
There are no parts you can repair by yourself. If the machine still can not run normally after the above
method, please contact the dealer or our company.
This device does not include accessories for repair usage; the repair should be carried out by an
authorized person or authorized service center.
• Keep the equipment in a dry storage condition.
• Do not smear the equipment with pigments.
• Calibration is recommended when using a new/other contra-angle handpiece.
• Replace the battery if it seems to be running out of power sooner than it should.
• Use soft cloth with ethanol to scrub the shell of the motor of the handpiece. Never use high
concentration ethanol or corrosive liquid.
• If you don’t usually use the product, please charge it often.
Can’t turning on
Troubles
Possible problems and treatment method
No display of charging
sign while charging
The auto-reverse does
not work
• The batteries are probably dead, please charge.
• Pressing time of key is too short, please press the power key for
over 1 second.
• The power key does not work, please contact local dealer or
manufacturer.
• The power adapter is not plugged in, please reconnect the power
adapter.
• The charging base is not properly placed in the charging seat.
• There is a foreign object on the bottom of the motor of the
handpiece or on the cradle, please clean up.
Batteries are weakened (lower remaining battery capacity), please
charge the batteries.
Troubleshooting Guide
Maintenance
10.1 Lubrication of contra-angle handpiece
• Lubrication should be limited to the contra-angle handpiece only. Apply lubricant spray after each
use and/or before sterilization.
• When you lubricate, please aim an injection nozzle at the rear part of the contra-angle and
lubricate for 2-3 seconds. • Do not lubricate the motor of the handpiece.
10.2 Cleaning motor of the handpiece
• When the motor of the handpiece becomes dirty, wipe it off with a cotton cloth moistened with
rubbing alcohol.
10.3 Cleaning charger
• When the charger becomes dirty, wipe it off with a cotton cloth moistened with surgical spirit.
10.4 Sterilization
• Sterilize the contra-angle handpiece only. For the sterilization method we recommend autoclaving.
• Please use a brush (do not use metal brush) to brush away the dirt attached to the surface, and then
wipe it with a clean soft cloth using surgical spirit.
• Please place the contra angle handpiece into a sterilization bag and seal it.
• The temperature should not exceed 135°C (eg 121°C for 20 minutes or 132°C for 15 minutes is ideal
for autoclaving).
• Keep the contra-angle handpiece in the autoclave pouch to keep it clean until you use it.
Enviromental Protection
11.1 Storage conditions
• Storage temperature range : -10°C - +50°C
• Relative humidity range : less than 95%
• Atmospheric pressure range : 50KPa - 106KPa
11.2 Environmental conditions
Do not expose to sun, rain, dust, corrosive gases or volatile chemicals keep the device in a ventilated
place.
Disinfection