After completing the key configuration and waiting for the screen to stop flashing, the program will
be autosaved, at this time press the Program Shift Key P to switch the program and the set value of the
previous page will be autosaved.
Please note:
Forward definition: when cutting file, move clockwise.
Press this key to increase or decrease the parameters of each item. When the current item is
flashing, press “+/-” key to change the parameter of the current item. While the file in contra-angle
is rotating, pressing “+/-” key can also change speed (SP) up and down. Please refer to the following
parameters while using.
The machine factory optional mini push-button 16: 1 Contra-angle Handpiece. When use, refer to the
following tables for the choice of parameter range:
Ratio
RPM
Torque (N/cm)
16:1
130-600
0.6-4.0
7
Mode Display
Display current mode, in the reciprocating function, the torque is not adjustable:
Positive auto-continuous (auto-reverse. if no loaded will be forward again}
Reverse reciprocating motion
Reverse automatic continuous (auto-forward, if not loaded will be reversed again
Forward reciprocating motion
Turn off automatic reverse(no automatic reverse)
Tum off automatic forward{no automatic forward)
Reverse the reciprocating motion, the speed can be adjusted from 200 to 400, reverse 150,
forward 30"(approximation)
Forward reciprocating motion, the speed can be adjusted from 200 to 400, forward 150",
reverse 30"(approximation)
Battery Capacity Display
The following is a symbol of the current battery capacity, the battery will flicker.
It indicates that the battery is full or will soon be full.
It indicates that the remaining capacity of batteries is 30%-80%.
It indicates that the remaining capacity of batteries is less than 30%.
It indicates that the battery is running low, please charge.
Reset Setting
While on standby(positive/negative movement), install Contra-angle Handpiece and do not give extra
load to it, then long-press Item Select key S when "CRL “ appears on the screen, press the motor switch
again to prompt"bee".At this time, the instrument will start to reset and then enter the lock screen
interfaces.
Note:
The use of the reset function is in situations when the setting is correctly installed or the machine
does not display or set the selected mode.
Liquid Crystal Display Panel Diagram
8