1. Setting up

NOTE: 

The sieving will not start if the lid is not on. 

• Start sieving by pressing the START button “>”. 

• Stop the sieving by pressing the STOP button “

Ƒ

“. 

Filter (Option) 

Electrostatic charge inside the devices is prevented by earthing the device via the protective 

conductor on the electrical connection.  Ensure the correct assembly of filter unit in order to 

guarantee sufficient earthing.   

NOTICE: 

Despite this, electrostatic charge separation may however still occur between the sample and 

receptacle wall inside the collecting receptacle depending on the sample property, flow speed and air 

humidity.  

Cleaning and service  

WARNING: 

Risk of a fatal electric shock 

- An electric shock can cause injuries in the form of burns and cardiac arrhythmia, respiratory arrest

or cardiac arrest. 

• Do not clean the machine under running water. Use only a cloth dampened with water. 

• Disconnect the power supply plug before cleaning the blender. 

¾

Cleaning 

We recommend ultrasonic baths for thorough, gentle and time-saving cleaning of your test sieves. 

Clean the air outlet channel regularly using a brush to remove any deposits.   

x

Vacuum cleaner – changing the vacuum cleaner bag

x

Change the vacuum cleaner bag or empty the dust container on your vacuum cleaner 
regularly as required. 

QAQC LAB 589 Rappahannock Drive White Stone Va 22578 TEL (804) 318-3686

Summary of Contents for Air Sizer 200

Page 1: ...1 Technical specifications Copyright by Endecotts Ltd 45 1 Technical specifications 6 Operating instructions Air Sizer 200 QAQC LAB 589 Rappahannock Drive White Stone Va 22578 TEL 804 318 3686...

Page 2: ...the sieve mesh constantly The Air Sizer 200 is specially designed for the dry sieving and the particle size determination of fine grained dry pourable and dispersed bulk materials The sieve holder is...

Page 3: ...actions 1 Amplitude N A Speed 5 to 55 rpm Time display digital 00 10 99 50 min Suitable for dry sieving yes Suitable for wet sieving no Sieve diameter 203 mm 8 premium air jet sieves Clamping device N...

Page 4: ...and the controls and indicators before operating this machine PLEXIGLASS LID FOR AIR JET SIEVING TEST SIEVE AIR OUTLET CHANNEL MAINS SWITCH CONTROL PANEL COVER BASE NOZZLE COMPARTMENT MAINS POWER CONN...

Page 5: ...icate START button ON 4 Start the machine by pressing the START button 5 7 and button reduces increases the sieving time 00 10 99 50 min 6 Stop the machine by pressing the STOP button 8 9 and button r...

Page 6: ...n change to an unknown degree Atmospheric humidity Maximum relative humidity 80 at temperatures up to 31 C with linear reduction down to 50 relative humidity at 40 C At high atmospheric humidity the e...

Page 7: ...perated Take Care the Air sizer 200 weighs 14 kg x 1 off Set up Instructions x 1 off Mains Cable x 1 off Air Sizer fitted x 1 off Plexiglas lid x Soft Faced Mallet x Test Sieving Manual Position the A...

Page 8: ...ith a diameter of 203mm 8 inches The range of mesh fineness extends from 10 m to approx 4mm x Place the sieve in the nozzle compartment x Fit the Plexiglas lid provided NOTE The Air Sizer 200 cannot b...

Page 9: ...ot be used as a production machine nor is it intended for continuous operation Note Defects in components due to liquids Penetration of liquids inside the housing x Components are damaged and the corr...

Page 10: ...er to life through electric shock x An electric shock can lead to burns and to cardiac arrhythmias or to respiratory arrest and cardiac arrest x The device may only be operated with plugs that have a...

Page 11: ...mbar sensor could be destroyed The air jet can be changed by turning the manual air jet setting 1 Air entry opening closed maximum air jet 2 Air entry opening open minimum air jet The difference in ai...

Page 12: ...on the sample property flow speed and air humidity Cleaning and service WARNING Risk of a fatal electric shock An electric shock can cause injuries in the form of burns and cardiac arrhythmia respirat...

Page 13: ...This device is essentially maintenance free if cleaned regularly E11 MOTOR FAULTY BLOCKED Service required E83 VACUUM TOO LOW Check whether the suction apparatus is connected the suction apparatus is...

Page 14: ...laboratory equipment Authorised for the compilation of technical documents Endecotts Ltd technical documentation The following records are held by Endecotts Ltd in the form of Technical Documentation...

Reviews: