Connection of the industrial vacuum cleaner 

 Connect your vacuum cleaners suction tube to the air outlet channel

 Insert the type F IEC C14 connector on the vacuum cleaner into the IEC C13 panel-mounted

outlet.  

The power for the industrial vacuum cleaner is supplied from the Air Sizer 200.

 

Negative Pressure 

Display shows the neg. Pressure........... (Range from 00 to 99 mbar) 

NOTE: 

If the neg. pressure is more than 99 mbar -> sensor could be destroyed 
The air jet can be changed by turning the manual air jet setting.  
[1] Air entry opening closed è maximum air jet 
[2] Air entry opening open è minimum air jet 
The difference in air pressure (differential pressure) between the air inlet and air outlet at a given 
moment is displayed. 

Operating Software  

The operating software version is displayed for 5 seconds having switched on the machine.

¾

Starting, Stopping 

• 

Switch the device on by pressing the on/off switch at the back. 

• 

set the sieving parameters you want. 

• 

Put the test sieve with the sample onto the nozzle compartment. 

• 

Place the lid on the sieve. 

QAQC LAB 589 Rappahannock Drive White Stone Va 22578 TEL (804) 318-3686

Summary of Contents for Air Sizer 200

Page 1: ...1 Technical specifications Copyright by Endecotts Ltd 45 1 Technical specifications 6 Operating instructions Air Sizer 200 QAQC LAB 589 Rappahannock Drive White Stone Va 22578 TEL 804 318 3686...

Page 2: ...the sieve mesh constantly The Air Sizer 200 is specially designed for the dry sieving and the particle size determination of fine grained dry pourable and dispersed bulk materials The sieve holder is...

Page 3: ...actions 1 Amplitude N A Speed 5 to 55 rpm Time display digital 00 10 99 50 min Suitable for dry sieving yes Suitable for wet sieving no Sieve diameter 203 mm 8 premium air jet sieves Clamping device N...

Page 4: ...and the controls and indicators before operating this machine PLEXIGLASS LID FOR AIR JET SIEVING TEST SIEVE AIR OUTLET CHANNEL MAINS SWITCH CONTROL PANEL COVER BASE NOZZLE COMPARTMENT MAINS POWER CONN...

Page 5: ...icate START button ON 4 Start the machine by pressing the START button 5 7 and button reduces increases the sieving time 00 10 99 50 min 6 Stop the machine by pressing the STOP button 8 9 and button r...

Page 6: ...n change to an unknown degree Atmospheric humidity Maximum relative humidity 80 at temperatures up to 31 C with linear reduction down to 50 relative humidity at 40 C At high atmospheric humidity the e...

Page 7: ...perated Take Care the Air sizer 200 weighs 14 kg x 1 off Set up Instructions x 1 off Mains Cable x 1 off Air Sizer fitted x 1 off Plexiglas lid x Soft Faced Mallet x Test Sieving Manual Position the A...

Page 8: ...ith a diameter of 203mm 8 inches The range of mesh fineness extends from 10 m to approx 4mm x Place the sieve in the nozzle compartment x Fit the Plexiglas lid provided NOTE The Air Sizer 200 cannot b...

Page 9: ...ot be used as a production machine nor is it intended for continuous operation Note Defects in components due to liquids Penetration of liquids inside the housing x Components are damaged and the corr...

Page 10: ...er to life through electric shock x An electric shock can lead to burns and to cardiac arrhythmias or to respiratory arrest and cardiac arrest x The device may only be operated with plugs that have a...

Page 11: ...mbar sensor could be destroyed The air jet can be changed by turning the manual air jet setting 1 Air entry opening closed maximum air jet 2 Air entry opening open minimum air jet The difference in ai...

Page 12: ...on the sample property flow speed and air humidity Cleaning and service WARNING Risk of a fatal electric shock An electric shock can cause injuries in the form of burns and cardiac arrhythmia respirat...