1. Technical specifications
Copyright ©
ŘŖŗŜ
by Endecotts Ltd.
55
1. Technical specifications
Connection of the industrial vacuum cleaner
•
Connect your vacuum cleaners suction tube to the air outlet channel
•
Insert the type F IEC C14 connector on the vacuum cleaner into the IEC C13 panel-mounted
outlet.
The power for the industrial vacuum cleaner is supplied from the Air Sizer 200.
Negative Pressure
Display shows the neg. Pressure........... (Range from 00 to 99 mbar)
NOTE:
If the neg. pressure is more than 99 mbar -> sensor could be destroyed
The air jet can be changed by turning the manual air jet setting.
[1] Air entry opening closed è maximum air jet
[2] Air entry opening open è minimum air jet
The difference in air pressure (differential pressure) between the air inlet and air outlet at a given
moment is displayed.
Operating Software
The operating software version is displayed for 5 seconds having switched on the machine.
¾
Starting, Stopping
•
Switch the device on by pressing the on/off switch at the back.
•
set the sieving parameters you want.
•
Put the test sieve with the sample onto the nozzle compartment.
•
Place the lid on the sieve.