EM3 AEM Monitor User’s Guide
8-1
8
Replacement Parts and Service
This section includes information on replacement parts and procedures for the AEM Monitor
and instructions for returning the monitor for service.
CAUTION
Refer all servicing to qualified personnel.
At the end of the monitor’s product life, dispose of per local regulations for electrical equipment.
Replacement Parts and Procedures
There are only two customer replaceable parts in the AEM Monitoring System. These parts
do not require removal of the monitor cover. All other service must be performed by the
Encision Service Center.
NOTICE
Removal of the monitor cover will void the AEM Monitor warranty.
Customer Replaceable Parts
Part Number
A.C. Power Cord
115V
00492-000
Fuse
5 × 20 mm, 1 Amp (250V)
04169-000
Replacing the Fuse
WARNING
Electric Shock Hazard. Disconnect the power cord before replacing parts.
NOTICE
Fuse replacement procedure is only for external fuses.
1. Release the fuse drawer by inserting a small flat-head screwdriver into the slot at the
bottom of the fuse drawer and pulling gently on the drawer. Slide the drawer out.
2. Remove the blown fuse from the drawer.
3. Replace the blown fuse with one of the same type and rating.
4. Slide the fuse drawer into its slot until it snaps into place.
Returning the AEM Monitor
Before returning the AEM Monitor to Encision, call the Encision Customer Service Center to
obtain a Return Material Authorization number (RMA) or call your Encision Representative for
assistance. If you are returning the monitor for service, clean the monitor, then ship it to the
Encision Service Center.
Obtaining a Return Material Authorization Number
Have the following information ready when you call:
hospital / clinic name / customer number
telephone number
department / address, city, state, and zip code
model number
serial number
description of the problem
Cleaning the Monitor
1. Follow the procedures approved by your institution or use a validated infection control
procedure.