System Setup
3-6
EM3 AEM Monitor User’s Guide
WARNING
Electric Shock Hazard. Do not activate the generator bipolar output when the Bipolar Adapter is
connected to the generator, but not connected to the AEM Monitor. Accessible pins of the adapter
may lead to shock or burns to surgical personnel.
Electric Shock Hazard. Do not connect wet accessories to the generator. Ensure that accessories
and adapters are correctly connected and that no metal is exposed.
Inspect accessories and cords for breaks, cracks, nicks or other damage before every use. Verify that
end of life indicators are not present. If any of these are present, do not use. Failure to observe this
precaution may result in injury or electrical shock to the patient or operating personnel.
7. Prepare the surgical instrument to be used for the procedure. Connect one end of the
bipolar instrument cord to the instrument and the other end to the Bipolar Instrument
receptacle on the AEM Monitor.
8. Adjust the volume of the clicks that indicate bipolar current flow using the Volume Control
buttons on the front panel of the monitor.
CAUTION
The Bipolar Current Indicator activation clicks when an accessory is active. Do not turn the volume
down below an audible level.
9. Turn on the electrosurgical generator. It should be in its normal operating state.
10. Adjust the bipolar output mode and power setting on the electrosurgical generator.
WARNING
Confirm proper power settings before proceeding with surgery. Use the lowest power setting possible
for the minimum time necessary to achieve the desired surgical effect.
11. After successful completion of these steps, the system is ready for operation. During
operation, the Bipolar Current Indicator illuminates, indicating the current flowing
between the instrument’s tines.
WARNING
Do not attempt to connect or disconnect any cable during power activation.