Controls, Indicators, and Receptacles
EM3 AEM Monitor User’s Guide
2-3
Instrument Insulation Fault
Indicator
The text “Replace Instrument & Cord” illuminates solid amber (for approximately 30
seconds) if an instrument insulation fault is detected for a hand control or foot control
instrument. See Section 6 for information on correcting fault conditions.
Foot Control Instrument Indicator
Illuminates solid amber until a cord, foot control instrument, and return electrode
are properly connected.
Illuminates solid green when a foot control instrument and a return electrode are
properly connected and ready for use.
Illuminates blinking green when a foot control instrument is activated.
Illuminates solid amber if a foot control instrument insulation fault or cord connection
fault is detected.
Instrument Cord Connection Fault
Indicator
(foot control)
Illuminates solid amber until a foot control instrument and cord are connected. If the
instrument cord is disconnected, the indicator illuminates blinking amber for 30 seconds
before returning to a solid amber illumination. See Section 6 for information on
correcting fault conditions.
Hand Control Instrument Receptacle
Connect the hand control instrument cord to this receptacle.
Foot Control Instrument Receptacle
Connect the foot control instrument cord to this receptacle.
13
12
11
15
14