Users’ Manual
EAN-26 Digital inclinometer operation
Page
|
13
4.1.10.1 EDIT readings
Readings can be verified or retaken using EDIT mode. Pressing Down button once enters application in
EDIT mode. Once DN button is pressed, Accept button becomes Edit button and small digits in green
colour below axis-A and axis-B will appear for showing stored log for the interval showing on screen. Lower
the probe into borehole and press DN button corresponding to probe position into borehole. Monitor stored
log and present log to verify the reading log. If any stored log found wrong, new reading can be stored by
pressing EDIT button. UP and DN button can be used to reach the desired probe position. Verify all
readings at different intervals up to which readings are doubtful. Application automatically exits from EDIT
mode when highest logged interval is reached. This feature saves lots of time and efforts.
4.1.10.2 Reading option settings
Pressing option button from readings screen (figure 5.20) opens options screen. Figure 5.21 shows the
reading options screen. Number of readings for averaging data for axis-A/B can be set using UP/DN
arrows. Taking average for displacement can be disabled by un-checking the tick box. Borehole log file
name contains pre-fix as site code and borehole code while suffix having reading start date and time stamp.
Comments can be entered for borehole log file. Advanced options menu can be opened by pressing
advanced button.
Figure 5.20