Electromagnetic Compatibility
Portable and mobile RF
communications equipment can
affect medical electrical
equipment. Avoid operating
AutoTouch Connect
TM
near
microwave ovens, wireless
routers, baby monitors or other
common household electronics
that operate using RF
transmission, including RFID
emitters. A minimum distance of
30 cm (12 inches) is
recommended.
Avoid operating AutoTouch
Connect near high magnetic or
other fields such as those around
MRI, CAT, or PET scanners.
AutoTouch Connect emits RF.
Electromagnetic Emissions
The autoinjector is intended for use in the electromagnetic
environment specified below. The user of the autoinjector should
ensure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic Environment -
Guidance
RF Emissions
(CISPR 11)
Group 1
The autoinjector uses RF
energy for its internal and
system interface functions. Its
RF emissions are very low and
are not likely to cause any
interference in nearby
electronic equipment.
RF Emissions
(CISPR 11)
Class B
The autoinjector is suitable for
use in all establishments,
including domestic
establishments and those
directly connected to the public
low-voltage power supply
network that supplies buildings
used for domestic purposes.
Harmonic
emissions
IEC 61000-3-2
Not
Applicable
Voltage
fluctuations/
flicker emissions
IEC 61000-3-3
Not
Applicable