![Enbrel AutoTouch Connect User Manual Download Page 23](http://html1.mh-extra.com/html/enbrel/autotouch-connect/autotouch-connect_user-manual_2405754023.webp)
AutoTouch Connect
TM
Reusable Autoinjector
Failure
Symbol
:
Red triangle with an
exclamation point, red status
button, and error sound.
Problem
:
Several errors have
occurred or AutoTouch Connect
has stopped working.
Action
:
Reset AutoTouch
Connect.
To reset AutoTouch Connect:
Hold AutoTouch Connect away
from skin and press the status
button to wake AutoTouch
Connect. The failure symbol
should begin blinking and a chime
should sound.
While the failure symbol is
blinking, press and hold door
button until all symbols are
temporarily displayed and the
status button blinks green. The
door button should be held for at
least 10 seconds. After a
successful reset, if an Enbrel Mini
®
single-dose prefilled cartridge is
still inside AutoTouch Connect,
remove it. Close the AutoTouch
Connect door. Then, for the next
injection, start by pressing the
door button to open Enbrel Mini
door.
If AutoTouch Connect does not
respond after three attempts to
reset, call
1-888-4ENBREL
(1-888-436-2735).