background image

Oxford

  Up

®

13 

The Oxford Deluxe Standing sling with clips is 
suitable for use with the Oxford Up. Various sizes 
are available depending on patient requirements.

NOTE:

 For full fitting instructions, please refer to the user guide supplied with each sling.

If your patient would benefit from additional reassurance and support, the Oxford Deluxe Standing 
sling with clips can be used in combination with the Up.

Place the sling around the patient just above the waist line and secure the waist belt around the 
patient using the clip. This strap need not be tight. It is there to hold the sling in position whilst 
attaching the sling to the Up. Ensure that the client’s arms are outside the sling.

6. Sling Compatibility & Attachment

Fitting Instructions:

Oxford

®

Up

13

English

6. Sling Compatibility & Attachment

The 

Oxford Deluxe Standing sling with clips

 is 

suitable for use with the Oxford Up. Various sizes 

are available depending on patient requirements.

WARNING

Joerns Healthcare recommends that slings be checked regularly and particularly before 

use for signs of fraying or damage. DO NOT use slings that are worn or damaged.

WARNING

OXFORD RECOMMENDS THE USE OF GENUINE OXFORD PARTS. Oxford sling, lift and 

transfer products are designed to be compatible with one another. For country specific 

guidance on sling use and compatibility, please refer to the sling label or contact your local 

market distributor or Joerns Healthcare.

WARNING

Refer to maximum weight capacity of product. Sling capacity is limited by the maximum 

capacity of the product it is used on.

Fitting Instructions:

NOTE: For full fitting instructions, please refer to the user guide supplied with each sling.

If your patient would benefit from additional reassurance and support, the Oxford Deluxe Standing 

sling with clips can be used in combination with the Up.

Place the sling around the patient just above the waist line and secure the waist belt around the 

patient using the clip. This strap need not be tight. It is there to hold the sling in position whilst 

attaching the sling to the Up. Ensure that the client’s arms are outside the sling.

Oxford

®

Up

13

English

6. Sling Compatibility & Attachment

The 

Oxford Deluxe Standing sling with clips

 is 

suitable for use with the Oxford Up. Various sizes 

are available depending on patient requirements.

WARNING

Joerns Healthcare recommends that slings be checked regularly and particularly before 

use for signs of fraying or damage. DO NOT use slings that are worn or damaged.

WARNING

OXFORD RECOMMENDS THE USE OF GENUINE OXFORD PARTS. Oxford sling, lift and 

transfer products are designed to be compatible with one another. For country specific 

guidance on sling use and compatibility, please refer to the sling label or contact your local 

market distributor or Joerns Healthcare.

WARNING

Refer to maximum weight capacity of product. Sling capacity is limited by the maximum 

capacity of the product it is used on.

Fitting Instructions:

NOTE: For full fitting instructions, please refer to the user guide supplied with each sling.

If your patient would benefit from additional reassurance and support, the Oxford Deluxe Standing 

sling with clips can be used in combination with the Up.

Place the sling around the patient just above the waist line and secure the waist belt around the 

patient using the clip. This strap need not be tight. It is there to hold the sling in position whilst 

attaching the sling to the Up. Ensure that the client’s arms are outside the sling.

WARNING

WARNING

WARNING

Joerns Healthcare recommends that slings be checked regularly and 
particularly before use for signs of fraying or damage. DO NOT use 
slings that are worn or damaged.

Refer to maximum weight capacity of product. Sling capacity is 
limited by the maximum capacity of the product it is used on.

OXFORD RECOMMENDS THE USE OF GENUINE OXFORD PARTS. 
Oxford sling, lift and transfer products are designed to be compatible 
with one another. For country specific guidance on sling use and 
compatibility, please refer to the sling label or contact your local 
market distributor or Joerns Healthcare.

!

!

!

Summary of Contents for Oxford Up

Page 1: ...S T R U C T I O N M A N U A L Product name to go here Product description to go here U S E R M A N U A L enablelifecare com au Oxford Up Ge Lees voor gebruik de Pour viter les blessures lisez le ma T...

Page 2: ...Oxford Up 2 Contents 1 Introduction 3 2 Assembly Commissioning Instructions 4 3 Safety Precautions 5 4 Operating Instructions 7 5 Maintenance Schedule 15 6 Technical Specifications 16 7 Warranty 17...

Page 3: ...arries the CE mark and complies with the following EC directives The CE mark Medical Device Directive MDR 2017 745 The Oxford Up is intended for use by an individual who has the ability to stand unass...

Page 4: ...ed assistance to lift the Oxford Up from its carton The Up requires no additional assembly once removed from its packaging it is ready for use in accordance with the guidelines and daily checks contai...

Page 5: ...suitable for the particular patient and is of the correct size and capacity NEVER use a sling belt which is frayed or damaged ALWAYS fit a sling belt according to the user instructions provided ALWAY...

Page 6: ...loaded device on a slope which exceeds 1 12 approximately 5 degrees Joerns Healthcare recommend a second helper is present when moving a patient on a slope DO NOT park a loaded device on ANY sloping s...

Page 7: ...n firmly on the integrated brake pedal To release the brakes simply manoeuvre the pedal back up Oxford Up 7 English 4 Operating Instructions 1 Swing away Seat Pads The seat pads are designed to swing...

Page 8: ...ior to using the Up Maintain a dialogue with the patient ensuring they understand what is happening and they are comfortable and at ease If a sling has been deemed necessary as part of the risk assess...

Page 9: ...heel the Up into position in front of the patient placing their feet on the footplate with their knees resting comfortably against the cushioned knee pad Once the patient is comfortably in position ap...

Page 10: ...urface are applied to avoid any inadvertent movement and ensure a safer more stable transfer Ask the patient to grasp the crossbar and prepare to raise themselves to a standing position Offer words of...

Page 11: ...ully with the transfer WARNING When transferring to or from a wheeled movable surface such as a commode wheelchair or bed ensure the brakes of that surface are applied to avoid any inadvertent movemen...

Page 12: ...glish Once the patient is standing rotate the swing away seat pads into the up position providing a barrier free space in behind WARNING When transferring to or from a wheeled movable surface such as...

Page 13: ...e clip This strap need not be tight It is there to hold the sling in position whilst attaching the sling to the Up Ensure that the client s arms are outside the sling Oxford Up 13 English 6 Sling Comp...

Page 14: ...nbalanced Continue with the transfer accoridng to the guidance in this document PLEASE NOTE 1 Attach Securi3 clips to the attachment pins as described in the illustration shown 2 Ensure all green safe...

Page 15: ...finish of the product ENSURE the product moves freely on its castors ENSURE both brakes are operational and engage correctly ENSURE the legs open and close correctly EXAMINE the sling attachment point...

Page 16: ...Leg Height 95mm Ground Clearance 15mm Overall Height Of Base Footplate 65mm Knee Pad Height To top of knee pad 465mm Turning Radius 1165mm Front Castors 75mm Rear Castors 100mm Outside this environme...

Page 17: ...be contacted should any problem arise If you are in any doubt as to where your product was purchased Joerns Healthcare can trace the supplier if you quote the serial number of the product All warrant...

Page 18: ...m au At Enable we source quality affordable and innovative products globally and have them available for fast delivery But we do more than just provide products We give the knowledge and support to he...

Reviews: