14.0
| Troubleshooting
If your unit does not seem to operate correctly, refer to the chart below to determine possible causes.
If none of these measures correct the problem, please contact a Customer Service Representative.
function properly.
Low Battery
indicator flash.
Recharge
battery pack
Recharge
battery pack
1. Check all control settings. Are they set to values prescribed
by your medical professional?
2. Are electrodes in proper position and adhering to the skin?
3. Check lead wires. Be sure all connectors are firmly sealed.
4. Replace cord set with another to check for broken wires.
None of LCD
indicators
illuminate.
1. Ensure battery pack is inserted in battery compartment
2. Ensure battery pack is matched with correct polarity as shown on battery and compartment
3. Ensure battery pack is not damaged
26
Manufacturer
Keep Dry
Single patient use
Do not put in regular trash
Caution: Read Instruction Manual
For Prescription use only