17
Functions
Descriptions
1/1
abrupt
When set at “1/1” with the frequency control set at 100Hz, the interference fre-
quency would be at 75Hz for 1 second, then shift abruptly to 155Hz for 1 second,
switch is set in the “1/1” mode.
C
(Continuous)
In the “C” (Continue) mode, there is no change in the pulse rate, When set at the
other modes, the interference frequency changes over time.
8/8
abrupt
(75Hz to 155Hz in the above example) is held for 8 seconds.
10/10
ramped
of rapidly. For example, when the frequency control was set at 100, the device will
sweep gradually from 75Hz to 155Hz over a 10 second period, then from 155Hz to
75Hz during the next 10 seconds.
10.0
| Interferential Treatment Functions Descriptions
NOTE:
In order to receive a true interferential treatment, channel 1 and 2 should be used in a cris-cross
pattern.With this setting, only channel 2 is used to increase the intensity.