ZERO/SPAN CALIBRATION CHECK
The zero calibration has been set at the factory by placing the instrument on an optical bench using the flange-to-
flange dimensions specified by the customer and recorded in the customers final test report. This zero value is
critical as it can offset smoke measurement plus or minus if not correctly set. The measurement reflector contains
an aperture that is chosen during the factory calibration and is fixed. The Transceiver is aligned with the beam
centered on the retro reflector and the electronics is adjusted to produce a zero opacity value.
The calibration zero reflector is then adjusted to provide a low value typically 1-2% and the span filter up scale
value is recorded. The values are entered in the
Quick Menu
under "Zero Cal Value" and " Span Cal Value".
When the system enters a manual or auto cal check cycle the current zero and span values are checked against the
stored values. If either the zero or span current value exceeds the stored value by +/- 2% opacity the fault system
will energize.
To verify initiate a manual cal cycle by pushing the
AUTO/MAN
key. The Alarm LED will light and the
Left/Right Bargraph will blink, this is normal because a manual cal is considered a fault condition. Pressing the
ACK
key will cause the ALARM LED and the bargraphs to stop blinking. The message display will read
MANUAL CAL ALARM. Push the
DIS
key to return to display opacity.
After the cal cycle is complete pushing the
ACK
key should clear any alarms. If no alarms are present the system is
working properly.
Summary of Contents for 1304
Page 45: ...44 Section 8 DRAWINGS ...
Page 46: ...45 EMS 1001 Mounting under 6 STACK ...
Page 47: ...46 EMS 1002 Mounting over 6 stack ...
Page 48: ...47 EMS 1024 Air purge Weather cover mech Installation ...
Page 49: ...48 EMS 1078 PANEL cut out for 1304 control unit ...
Page 50: ...49 EMS 1079 1304 control unit dimensions ...
Page 51: ...50 EMS 1077 Terminal identification 1304 control unit ...
Page 52: ...51 EMS 1131 1304 with recorder system wiring ...
Page 54: ...53 EMS1101 Sht 1 of 3 Dual Blower system wiring ...
Page 55: ...54 EMS 1101 Sht 2 of 3 Single Blower system wiring ...
Page 56: ...55 EMS 1101 Sht 3 of 3 Without Blower system wiring ...