26
Control unit manual controls
ACK
-
Acknowledges fault alarms and displays the fault message.
DIS
-
Changes display each time it is pressed; Opacity %, Average Opacity, dust/zero shift, hours to next cal.
AUTO/MAN
-
initiates a maintenance cal cycle and resets calibration cycle timer clock.
Alarm set point adjustments
Under the flip down door, press the
EDIT (ESC)
button (the red Edit
(ESC) LED will light). The message display will read QUICK MENU.
Press the
ENTER
key, then the
EDIT
key 1 time until the Display
reads ALARM SET POINT. Press the
ENTER
key twice, the Edit
Value Light will flash. Press and hold the
EDIT
or
EDIT
key
until the desired set point in % opacity value is displayed, i.e. 20.00 is
20.00% opacity. Press the
EDIT
key 4 times to return to normal
operation.
Alarm Time Delay Adjustment
Under the flip down door, press the
EDIT (ESC)
button (the red Edit
(ESC) LED will light). The message display will read QUICK MENU.
Press the
ENTER
key, then the
EDIT
key 2 times until the Display
reads ALARM TIME DELAY. Press the
ENTER
key twice, the Edit
Value Light will flash. Press and hold the EDIT
or
EDIT
key until
the desired time in seconds is displayed, i.e. 15 is 15 seconds delay. Press
the
EDIT
key 4 times to return to normal operation.
(ESC)
EDIT
ENTER
EDIT
EDIT
REM
LOC
AUTO
MAN
DIS
ACK
+19999
MESSAGE
PV
SP
OUT
EDIT
Keys
Hinged
Cover
“Edit
Value”
Light
“Scroll
Menu”
Light
Moves
Message
Cursor
Summary of Contents for 1304
Page 45: ...44 Section 8 DRAWINGS ...
Page 46: ...45 EMS 1001 Mounting under 6 STACK ...
Page 47: ...46 EMS 1002 Mounting over 6 stack ...
Page 48: ...47 EMS 1024 Air purge Weather cover mech Installation ...
Page 49: ...48 EMS 1078 PANEL cut out for 1304 control unit ...
Page 50: ...49 EMS 1079 1304 control unit dimensions ...
Page 51: ...50 EMS 1077 Terminal identification 1304 control unit ...
Page 52: ...51 EMS 1131 1304 with recorder system wiring ...
Page 54: ...53 EMS1101 Sht 1 of 3 Dual Blower system wiring ...
Page 55: ...54 EMS 1101 Sht 2 of 3 Single Blower system wiring ...
Page 56: ...55 EMS 1101 Sht 3 of 3 Without Blower system wiring ...