Phoenix User’s Manual
28
11. Clinician Only Section
This section is only intended for clinicians and the functions described in this section should
not be operated by the patient.
11.1 Compliance Monitoring
The Empi Phoenix device is equipped with a compliance monitoring system. This allows you
to monitor the usage of the device by your patient between visits.
To access the compliance screen, simultaneously and
continuously push on the Program Decrease
, Left
Decrease Intensity
and Right Decrease Intensity
buttons while you are in the Program Selection mode.
In the compliance screen, you will see the total time spent
in each of the programs by switching between the different
programs (for this, use the P- and P+ keys).
P3 12:36
HOURS
MINUTES
Note:
Timer indicates hours and minutes instead of minutes and seconds when you are in the
Compliance screen.
Compliance counters reset
To reset the compliance counters, press continuously and simultaneously for more than 3
seconds the Program Decrease
, Left and right Decrease intensity (same key combination
as to access the compliance screen).
To exit the compliance screen, switch Off the device by pressing the ON/OFF
switch.