4.
The sequence indicator will indicate the number of sequences for each program. The
program timer indicates the total duration of the program. Once you have selected the
desired program, simply increase the intensities of the channels you want using the
Increase Intensity buttons. For intensity strength information, refer to 6.6.
5.
The intensity of each channel will be displayed both numerically
and with a bar graph. Note that intensities of Endurance (P1)
and Strength (P2) programs are expressed in Energy levels (max
520) while intensities of TENS (P3) and Edema (P4) programs are
expressed in mA (max 100 mA).
7.4.4 Pause function
During a treatment, you can pause the treatment at any time by pressing the ON/OFF
button once. In this case, both intensities will drop to 0, the timer will stop, and the timer
indicator will blink.
To resume the treatment, increase the intensities to the desired level. This will automatically
resume the timer.
When you are in the Pause mode, you can also switch off the unit by pressing the ON/OFF
button a second time.
7.4.5 End of treatment
For all treatments except P3 (TENS), the timer will automatically finish the treatment. When the
treatment is finished, intensities are automatically reduced to 0 and the timer indicator will flash
with 00:00.
For the TENS (P3) program, the timer is unlimited, meaning that it does not automatically
shut off after a set time. Stop the treatment after the time recommended by your healthcare
provider by pressing the ON/OFF button.
• Turn off the device by pressing the ON/OFF button.
Disconnect the electrodes and the lead wires as indicated in section 7.3. Electrodes no longer
fit for use can be disposed of with the normal domestic waste.
7.4.6 Use of the hand switch
(
optional)
The hand switch connects to the unit on the dedicated port:
The hand switch can be used to manually change the cycles of contractions/active‐rest or turn
the stimulation On and OFF.
• In a constant mode (P3, P4, warm up and cool down of P1 and P2), pushing the button
will activate the stimulation while releasing it will bring all intensities to 0.
• In Work mode, pushing and releasing the button, respectively, will toggle between
contraction and active-rest phases.