Figure 1-2: Accessing Modbus ID for Ports A and B
4. Based on the information contained in the connection properties
window, enter the values for the following parameters in the
Summary table of communications/output settings
Port A and B: Modbus ID = _____(enter Comms Address).
5. Exit MeterLink and disconnect your PC from the meter electronics.
The remaining information required before doing the retrofit
upgrade, is based on the wiring terminations and the settings of
various switches on the boards that comprise the electronics.
6. Power down the electronics.
CAUTION
Equipment damage
Remove power from the meter electronics.
Failure to remove power from the meter electronics may result in
damage to the equipment.
7. Remove the end caps from the upper electronics housing of the
meter to access the electronic circuit boards. See
for
board layout of the 3400 stacks. See
electronics for board layout.
Upgrade kit instructions
February 2022
8
3410 Series Electronics Upgrade kit instructions
Summary of Contents for Rosemount Ultrasonic 3410 Series
Page 2: ...Upgrade kit instructions February 2022 2 3410 Series Electronics Upgrade kit instructions ...
Page 4: ...Upgrade kit instructions February 2022 4 3410 Series Electronics Upgrade kit instructions ...
Page 55: ...February 2022 Upgrade kit instructions Upgrade kit instructions 55 ...
Page 71: ...February 2022 Upgrade kit instructions Upgrade kit instructions 71 ...
Page 73: ......
Page 74: ... ...
Page 75: ...WARNING WARNING 3 4 NPT 3 4 NPT US C LISTED 90Y1 TELEMETERING EQUIPMENT FOR HAZ LOC ...
Page 76: ... ...
Page 77: ...February 2022 Upgrade kit instructions Upgrade kit instructions 73 ...
Page 78: ...Upgrade kit instructions February 2022 74 3410 Series Electronics Upgrade kit instructions ...
Page 79: ...February 2022 Upgrade kit instructions Upgrade kit instructions 75 ...