2.4
Purging the carrier gas lines
Use this procedure to purge the gas chromatograph's gas lines.
Refer to
Procedure
1. Close all valves and tighten all fittings.
2. Run tubing to the GC, but do not connect.
3. Back off pressure regulator (turn counterclockwise) fully.
Refer to
Figure 2-4: Carrier Gas Valve Configuration
4. Open cylinder valve for Carrier Cylinder 1.
The pressure indicator will read the cylinder pressure.
5. Open the shut-off valve attached to the carrier regulator.
6. Regulate pressure out of the cylinder to 20 psig; then close the cylinder valve.
7. Open V-1 (bleed valve) and let the carrier gas bleed to atmosphere until both
gauges read 0 psig and then close V-1.
8. Repeat
twice to purge the line to V-2.
9. Purge the line to V-3 by repeating
through
, but this time, use bleed
valve V-4 and Carrier Cylinder 2.
Retrofit Instructions
Installation
SK-07781
September 2019
Rosemount Model 500 Helium to Hydrogen Gas Conversion Kit
13