NOTICE
Recalibration is required whenever the electronic cards or sensing cell are replaced.
8.3.2
Replace transmitter board
The transmitter board is not repairable and must be replaced if any component fails.
See
Figure 8-1
on page 66 for item number references. The transmitter board is available
as a standalone kit or with the plastic cover and mounting plate.
WARNING
ELECTRIC SHOCK
Disconnect and lock out power before working on any electrical components.
CAUTION
MEASUREMENT ERRORS
Failure to resynchronize the calibration parameters with a Rosemount 6888Xi after
replacing the transmitter board or recalibrating the instrument may cause an inaccurate
O
2
measurement. When the transmitter board is replaced and the calibration parameters
are not synchronized with the Rosemount 6888Xi, the parameters in the Rosemount
6888Xi will be used as default to calculate the O
2
measurement.
Instruction Manual
Maintenance and service
00809-0100-4891
July 2018
Instruction Manual
65
Summary of Contents for Rosemount 6888C
Page 10: ...General information Instruction Manual July 2018 00809 0100 4891 10 6888C ...
Page 14: ...Typical system package Instruction Manual July 2018 00809 0100 4891 14 6888C ...
Page 20: ...Specifications Instruction Manual July 2018 00809 0100 4891 20 6888C ...
Page 46: ...Install Instruction Manual July 2018 00809 0100 4891 46 6888C ...
Page 55: ...Instruction Manual Calibration 00809 0100 4891 July 2018 Instruction Manual 55 ...
Page 56: ...Calibration Instruction Manual July 2018 00809 0100 4891 56 6888C ...
Page 61: ...Instruction Manual HART menu trees 00809 0100 4891 July 2018 Instruction Manual 61 ...
Page 62: ...HART menu trees Instruction Manual July 2018 00809 0100 4891 62 6888C ...
Page 78: ...Maintenance and service Instruction Manual July 2018 00809 0100 4891 78 6888C ...
Page 82: ...Service support Instruction Manual July 2018 00809 0100 4891 82 6888C ...
Page 87: ...Instruction Manual 00809 0100 4891 July 2018 Instruction Manual 87 ...