Figure 6-1: Calibration Defaults
Probe #1
Detailed
setup
Calibration
Auto calibration
Analog
output
Tolerance
check
Calibration
recommended
diagnostic
Calibration
acknowledge
Cal gas
#1
Cal gas
#2
Cal gas
flow
time
Purge time
Dif
fuser
Defaults
T
rack cal gases
Of
f
Of
f
Of
f
0.4% O
2
8% O
2
300 seconds
300 seconds
Plugg
ed dif
fuser
Auto cal advance
Rate
On
Enable
autocal
Start on cal recommended
Cal interval
Next cal time
No – Std. housing
Y
es – Autocal
housing
No
672 hours
(1 week)
V
ariable
9999
= manually
initiated
autocal
Calibration
Instruction Manual
July 2018
00809-0100-4891
54
6888C
Summary of Contents for Rosemount 6888C
Page 10: ...General information Instruction Manual July 2018 00809 0100 4891 10 6888C ...
Page 14: ...Typical system package Instruction Manual July 2018 00809 0100 4891 14 6888C ...
Page 20: ...Specifications Instruction Manual July 2018 00809 0100 4891 20 6888C ...
Page 46: ...Install Instruction Manual July 2018 00809 0100 4891 46 6888C ...
Page 55: ...Instruction Manual Calibration 00809 0100 4891 July 2018 Instruction Manual 55 ...
Page 56: ...Calibration Instruction Manual July 2018 00809 0100 4891 56 6888C ...
Page 61: ...Instruction Manual HART menu trees 00809 0100 4891 July 2018 Instruction Manual 61 ...
Page 62: ...HART menu trees Instruction Manual July 2018 00809 0100 4891 62 6888C ...
Page 78: ...Maintenance and service Instruction Manual July 2018 00809 0100 4891 78 6888C ...
Page 82: ...Service support Instruction Manual July 2018 00809 0100 4891 82 6888C ...
Page 87: ...Instruction Manual 00809 0100 4891 July 2018 Instruction Manual 87 ...