4
Configuration
4.1
User interface
Figure 4-1: Features of the User Interface
1 1 2
6 7 8
3 4 5
CAL
TEST
MENU
LED
Display
Table 4-1: LEDs
Green
Relay is energized
Yellow
Relay is de-energized
Red
Maintenance (flashing) or error (not flashing)
4.2
Powering up the first time (calibration)
Calibration automatically starts when the Rosemount 2555 is powered up
for the first time. If the level switch is powered off and then on again, this
calibration procedure is not repeated when starting up.
Prerequisites
•
The level switch must be correctly mounted and wired.
•
The level of the solids material must be below the probe.
Procedure
1.
Ensure the level of the solids material is not covering the probe.
Quick Start Guide
October 2020
20
Quick Start Guide