Table 4-5: Auto Recalibration Menu (Advanced Menu)
Display
Description
Menu item
ON
Auto recalibration to uncovered probe
.
It is possible to commission an already filled silo
(covered probe). A proper calibration is not
possible with covered probe. A solution is to do
an auto calibration as soon as the silo becomes
empty (uncovered probe).
To do this, set Auto recalibration to
ON
and do a
push-button calibration with a covered probe
(press and hold the
CAL
button for 3 seconds).
The level switch will recalibrate (as an uncovered
probe) when the measured capacitance is 50% of
the sensitivity setting (menu item
D
) for more
than two minutes.
Do not set to
ON
if excessive solids material
build-up is present, since this build-up may
decrease the measured capacitance and cause
an incorrect calibration.
(1)
Menu item "F" is not valid, and will not appear on the display, if manual
calibration (Menu item "G") is set to "ON".
(2)
Factory default setting.
4.9.2
Manual calibration
Note
The LED will be red and flashing while the menu is displayed.
Table 4-6: Manual calibration menu (Advanced menu)
Display
Description
Menu item
G.
ON
Manual calibration ON/OFF
.
If set to
ON
:
•
Menu items
H
,
K
and
L
appear.
•
Menu items
D
(Quick-start menu) and
F
(Auto re-calibration) are no longer valid and
are hidden.
•
Push-button calibration is not possible (if
CAL
button is pressed, the display shows
G.ON
).
October 2020
Quick Start Guide
Quick Start Guide
31