Maximum operating temperature and pressure
(continued)
Table 1-2:
Body material option
Maximum temperature
Maximum pressure
Maximum pressure (for
CRN registration only)
-03 (Glass-filled PEEK (high
temperature))
392 °F (200 °C)
295 psig (2135 kPa)
220 psig (1618 kPa [abs])
-04 (Glass-filled Tefzel)
248 °F (120 °C)
200 psig (1480 kPa)
150 psig (1135 kPa [abs])
-05 (Unfilled Tefzel)
248 °F (120 °C)
200 psig (1480 kPa)
150 psig (1135 kPa [abs])
Insertion adapter specifications
Table 1-3:
Adapter
part num-
ber
Sensor com-
patibility
Process con-
nection
Wetted ma-
terials
Maximum
tempera-
ture
Maximum
pressure
Maximum
pressure
(for CRN
registration
only)
Weight /
shipping
weight
23242-02
For use with
option -21
1-1/2 in.
MNPT
316 stainless
steel, glass-
filled PEEK,
Viton
392 ° F
(200 °C)
295 psig
(2135 kPa)
220 psig
(1618 kPa
[abs])
3 lb / 4 lb
(1.5 kg / 2.0
kg)
23242-03
For use with
option -20
1-1/2 in.
MNPT
316 stainless
steel, glass-
filled PEEK,
Viton
392 °F
(200 °C)
295 psig
(2135 kPa)
220 psig
(1618 kPa
[abs])
3 lb / 4 lb
(1.5 kg / 2.0
kg)
2001990
For use with
option -21
2 in. MNPT
CPVC
Viton
100 °F
(38 °C)
100 psig
(791 kPa
[abs])
N/A
1 lb / 2 lb
(0.5 kg / 1.0
kg)
185 °F (85 °
C)
45 psig (412
kPa [abs])
Retraction assembly specfications
Table 1-4:
Specification
Description
Sensor compatibility
The retraction assemblies are used with 228 - [ ]-20-54-62
only
Wetted materials
315 stainless steel, ethylene polypropylene (EP), unfilled
Teflon, carbon-filled Teflon
Process connection
1 - 1/2 in. MNPT
Maximum operating condition
392 °F (200 °C), 295 psig (2135 kPa [abs])
Description and specifications
2
Reference Manual
Summary of Contents for Rosemount 228
Page 1: ...Reference Manual LIQ MAN 228 Rev V April 2018 Rosemount 228 Toroidal Conductivity Sensors ...
Page 2: ......
Page 4: ......
Page 6: ...Contents ii Reference Manual ...
Page 26: ...Wiring diagram for Rosemount 1056 and 56 Transmitters Figure 3 2 Wiring 20 Reference Manual ...
Page 34: ...Calibration 28 Reference Manual ...
Page 36: ...Maintaining and Troubleshooting 30 Reference Manual ...
Page 38: ...Accessories 32 Reference Manual ...
Page 40: ...Return of Materials 34 Reference Manual ...
Page 41: ...Return of Materials Rosemount 228 Sensor 35 ...
Page 42: ...Return of Materials 36 Reference Manual ...