
Measuring standard temperature
Figure 4-3:
Be sure air bubbles are not adhering to the sensor. An air bubble trapped in the
toroid opening has a particularly severe effect on the reading.
5.
Turn off automatic temperature compensation in the transmitter.
This eliminates error in the cell constant.
6.
Adjust the transmitter reading to match the conductivity of the standard.
4.3
Calibrating against a Referee - in-Process
Prerequisites
If possible, adjust the conductivity of the process liquid so that it is near the midpoint of
the operating range. If this is not possible, adjust the conductivity so that it is at least 5000
µS/cm.
Turn off automatic temperature compensation in the transmitter. This eliminates error in
the cell constant.
Procedure
1.
Connect the process and referee sensors in a series.
Keep tubing runs between the sensors short and adjust the sample flow to as high a
rate as possible. Short tubing runs and high flow ensure that the temperature of the
liquid does not change as it flows from one sensor to another.
2.
Allow the process liquid to flow through both sensors.
Orient the referee sensor so that the air bubbles always have an easy escape path
and cannot get trapped. Tap and hold the flow cell in different positions to allow
bubbles to escape.
Wait for readings to stabilize before starting the calibration.
Calibration
Rosemount 228 Sensor
25
Summary of Contents for Rosemount 228
Page 1: ...Reference Manual LIQ MAN 228 Rev V April 2018 Rosemount 228 Toroidal Conductivity Sensors ...
Page 2: ......
Page 4: ......
Page 6: ...Contents ii Reference Manual ...
Page 26: ...Wiring diagram for Rosemount 1056 and 56 Transmitters Figure 3 2 Wiring 20 Reference Manual ...
Page 34: ...Calibration 28 Reference Manual ...
Page 36: ...Maintaining and Troubleshooting 30 Reference Manual ...
Page 38: ...Accessories 32 Reference Manual ...
Page 40: ...Return of Materials 34 Reference Manual ...
Page 41: ...Return of Materials Rosemount 228 Sensor 35 ...
Page 42: ...Return of Materials 36 Reference Manual ...