If wall effects are absent in the process installation, use a sufficiently large container
for calibration to ensure that wall effects are absent. To check for wall effects, fill the
container with solution and place the sensor in the center, submerged at least 3/4 of
the way up the stem. Note the reading. Then move the sensor small distances from
the center and note the reading in each position. The readings should not change.
If wall effects are present, be sure the vessel used for calibration has exactly the
same dimensions as the process piping. Also ensure that the orientation of the
sensor with respect to the piping is exactly the same in the process and calibration
vessels. See
Figure 4-2
.
Calibration installation orientation
Figure 4-2:
3.
Rinse the sensor with water.
4.
Immerse the rinsed sensor in the standard solution.
Use a good quality calibrated thermometer to measure the temperature of the
standard solution. The thermometer error should be less than ±1 °C. Allow adequate
time for the solution and sensor to reach thermal equilibrium. If the sensor is being
calibrated in an open beaker, keep the thermometer far enough away from the
sensor so it does not introduce wall effects. If the sensor is being calibrated in a pipe
tee or similar vessel, it is impractical to place the thermometer in the standard
solution. Instead, put the thermometer in a beaker of water placed next to the
callibration vessel. Let both come to thermal equilibrium with the ambient air before
continuing calibration. See
Figure 4-3
.
Calibration
24
Reference Manual
Summary of Contents for Rosemount 228
Page 1: ...Reference Manual LIQ MAN 228 Rev V April 2018 Rosemount 228 Toroidal Conductivity Sensors ...
Page 2: ......
Page 4: ......
Page 6: ...Contents ii Reference Manual ...
Page 26: ...Wiring diagram for Rosemount 1056 and 56 Transmitters Figure 3 2 Wiring 20 Reference Manual ...
Page 34: ...Calibration 28 Reference Manual ...
Page 36: ...Maintaining and Troubleshooting 30 Reference Manual ...
Page 38: ...Accessories 32 Reference Manual ...
Page 40: ...Return of Materials 34 Reference Manual ...
Page 41: ...Return of Materials Rosemount 228 Sensor 35 ...
Page 42: ...Return of Materials 36 Reference Manual ...