Install
2.1
Installing the sensor
Keep at least 1 in. (25 cm) between the sensor and the pipe wall. If clearance is too small,
calibrate the sensor in place. Ensure that the sensor is completely submerged in liquid.
Mounting the sensor in a vertical pipe run with the flow from bottom to top is best. If the
sensor must be mounted in a horizontal pipe run, orient the sensor in the 3 o-clock or 9
o'clock position.
Rosemount 228 dimensional drawing
Figure 2-1:
Install
Rosemount 228 Sensor
5