7.3.6
Measured level is occasionally unstable
Figure 7-6: Symptom
A
B
C
D
A. Level
B. Time
C. Actual level
D. Reported level
Possible cause
The product surface is close to a suppressed false echo.
Recommended actions
•
If possible, remove the disturbing object.
Related information
Reference Manual
Service and troubleshooting
00809-0200-4480
May 2021
Reference Manual
53