7.2.13
Maintenance required - Cleaning
Alarm classification
Device status
Maintenance required
Class
Notification
Possible cause
Product build-up on the antenna.
Recommended actions
Clean the antenna.
Use a damp cloth and a mild cleaning agent suitable for the media and wetted parts of
the transmitter.
Note
Take care not to scratch the PTFE sealing surfaces.
Related information
Material exposed to tank atmosphere
Components of the transmitter
7.2.14
Max EEPROM write cycles expired – Check master and
restart the device
Alarm classification
Device status
Maintenance required
Class
Warning
Possible cause
Frequent writing of parameters from master to device.
Recommended actions
1. Check the master configuration.
2. Restart the device.
3. If error persists, contact your local Emerson representative.
Related information
Service and troubleshooting
Reference Manual
May 2021
00809-0200-4480
46