Calibration procedure
Reference Manual
June 2016
MAN-0076, Revision 09
26
Calibration procedure
The “Zero” calibration option is selected if the sensor is only being zeroed (this not a complete
calibration)
It does not require the application of span gas, as only the sensor’s zero point is adjusted. Ensure that
no contaminants are present, if the surrounding air is to be used for Zeroing. If Zero calibration is
needed, at step 7 above, select ‘Zero’ using switch 3).