To mount the core processor:
1. Identify the components shown in
. For dimensions, see
2. If desired, reorient the core processor housing on the bracket.
a. Loosen each of the four cap screws (4 mm).
b. Rotate the bracket so that the core processor is oriented as desired.
c. Tighten the cap screws, torquing to 30 to 38 in-lbs (3 to 4 N-m).
3. Attach the mounting bracket to an instrument pole or wall. For pipe mount, two
user-supplied U-bolts are required. Contact Micro Motion to obtain a pipe-mount
installation kit if required.
Figure 3-2 Remote core processor components
End-cap (to 9-wire ground screw)
Mounting bracket
Core processor cover
Core processor housing
Conduit opening
for 4-wire cable
Conduit opening
for 9-wire cable
4 X Cap screws (4 mm)