124
Micro Motion Series 1000 and Series 2000 Transmitters
Measurement Performance
Figure 10-17
Temperature calibration – ProLink II
Enter temperature of low-
temperature fluid
Temperature Offset calibration
Do Cal
Wait until sensor achieves
thermal equilibrium
Fill sensor with low-
temperature fluid
Calibration in Progress
light turns green
Calibration in Progress
light turns red
Close
Enter temperature of high-
temperature fluid
Temperature Slope calibration
Do Cal
Wait until sensor achieves
thermal equilibrium
Fill sensor with high-
temperature fluid
Calibration in Progress
light turns green
Calibration in Progress
light turns red
Close
Done
ProLink Menu >
Calibration >
Temp offset cal
ProLink Menu >
Calibration >
Temp slope cal
Summary of Contents for MICRO MOTION 1500
Page 10: ...viii Micro Motion Series 1000 and Series 2000 Transmitters ...
Page 18: ...8 Micro Motion Series 1000 and Series 2000 Transmitters ...
Page 24: ...14 Micro Motion Series 1000 and Series 2000 Transmitters ...
Page 38: ...28 Micro Motion Series 1000 and Series 2000 Transmitters ...
Page 44: ...34 Micro Motion Series 1000 and Series 2000 Transmitters ...
Page 64: ...54 Micro Motion Series 1000 and Series 2000 Transmitters ...
Page 74: ...64 Micro Motion Series 1000 and Series 2000 Transmitters ...
Page 106: ...96 Micro Motion Series 1000 and Series 2000 Transmitters ...
Page 172: ...162 Micro Motion Series 1000 and Series 2000 Transmitters ...
Page 178: ...168 Micro Motion Series 1000 and Series 2000 Transmitters ...
Page 190: ...180 Micro Motion Series 1000 and Series 2000 Transmitters ...
Page 268: ...258 Micro Motion Series 1000 and Series 2000 Transmitters ...
Page 280: ...270 Micro Motion Series 1000 and Series 2000 Transmitters ...
Page 281: ......