
Configuration and Use Manual
109
Measurement Performance
Meas
ure
m
ent P
e
rf
ormance
T
rou
b
leshoo
tin
g
Custod
y T
rans
fer
Compensa
tion
10.3.4
Reading and interpreting meter verification test results
Pass/Fail/Abort
When the meter verification test is completed, the result is reported as Pass, Fail or Caution
(depending on whether you are using the display, the Communicator, or ProLink II), or Abort:
•
Pass
– The test result is within the specification uncertainty limit. In other words, the stiffness
of the left and right pickoffs match the factory values plus or minus the specification uncertain
limit. If transmitter zero and configuration match factory values, the sensor will meet factory
specifications for flow and density measurement. It is expected that meters will pass meter
verification every time the test is run.
•
Fail/Caution
– The test result is not within the specification uncertainty limit. Micro Motion
recommends that you immediately repeat the meter verification test. If you were using Smart
Meter Verification, with outputs set to Continue Measurement, change the setting to Last
Measured Value or Fault.
-
If the meter passes the second test, the first Fail/Caution result can be ignored.
-
If the meter fails the second test, the flow tubes may be damaged. Use your process
knowledge to determine the possibilities for damage and the appropriate actions for each.
These actions might include removing the meter from service and physically inspecting
the tubes. At minimum, you should perform a flow validation and a density calibration.
•
Abort
– A problem occurred with the meter verification test (e.g., process instability). Abort
codes are listed and defined in Table 10-3, and suggested actions are provided for each code.
Table 10-3
Meter verification abort codes
Abort code
Description
Suggested action
1
User-initiated abort
None required. Wait for 15 seconds before
starting another test.
3
Frequency drift
Ensure that temperature, flow, and density are
stable, and rerun the test.
5
High drive gain
Ensure that flow is stable, minimize entrained
gas, and rerun the test.
8
Unstable flow
Review the suggestions for stable flow in
Section 10.3.1 and rerun the test.
13
No factory reference data for meter verification
test performed on air
Contact Micro Motion customer service and
provide the abort code.
14
No factory reference data for meter verification
test performed on water
Contact Micro Motion customer service and
provide the abort code.
15
No configuration data for meter verification
Contact Micro Motion customer service and
provide the abort code.
Other
General abort.
Repeat the test. If the test aborts again, contact
Micro Motion customer service and provide the
abort code.
Summary of Contents for MICRO MOTION 1500
Page 10: ...viii Micro Motion Series 1000 and Series 2000 Transmitters ...
Page 18: ...8 Micro Motion Series 1000 and Series 2000 Transmitters ...
Page 24: ...14 Micro Motion Series 1000 and Series 2000 Transmitters ...
Page 38: ...28 Micro Motion Series 1000 and Series 2000 Transmitters ...
Page 44: ...34 Micro Motion Series 1000 and Series 2000 Transmitters ...
Page 64: ...54 Micro Motion Series 1000 and Series 2000 Transmitters ...
Page 74: ...64 Micro Motion Series 1000 and Series 2000 Transmitters ...
Page 106: ...96 Micro Motion Series 1000 and Series 2000 Transmitters ...
Page 172: ...162 Micro Motion Series 1000 and Series 2000 Transmitters ...
Page 178: ...168 Micro Motion Series 1000 and Series 2000 Transmitters ...
Page 190: ...180 Micro Motion Series 1000 and Series 2000 Transmitters ...
Page 268: ...258 Micro Motion Series 1000 and Series 2000 Transmitters ...
Page 280: ...270 Micro Motion Series 1000 and Series 2000 Transmitters ...
Page 281: ......