102
Micro Motion Series 1000 and Series 2000 Transmitters
Measurement Performance
Note: If you start a meter verification test from ProLink II or the Communicator, the transmitter
display shows the following message:
SENSOR
VERFY/
x
%
Figure 10-1
Meter verification test, original version – ProLink II
Verify configuration
parameters
Tools >
Meter Verification >
Structural Integrity Method
View previous test data
Next
Enter optional test data
Initialize and start meter
verification
Next
Abort
Next
Fault
configuration
Hold last
value
Progress bar shows
test in progress
Next
Finish
(2)
Graph of results
Rerun
test?
Yes
Pass
Fail
No
Start
Back
(1)
View report (option to print
or save)
Back
Abort
(1) If the graph was viewed at the beginning of the
procedure, clicking Back here will return to the
beginning of the procedure (along the dotted
line).
(2) The results of the meter verification test are not
saved until Finish is clicked.
Summary of Contents for MICRO MOTION 1500
Page 10: ...viii Micro Motion Series 1000 and Series 2000 Transmitters ...
Page 18: ...8 Micro Motion Series 1000 and Series 2000 Transmitters ...
Page 24: ...14 Micro Motion Series 1000 and Series 2000 Transmitters ...
Page 38: ...28 Micro Motion Series 1000 and Series 2000 Transmitters ...
Page 44: ...34 Micro Motion Series 1000 and Series 2000 Transmitters ...
Page 64: ...54 Micro Motion Series 1000 and Series 2000 Transmitters ...
Page 74: ...64 Micro Motion Series 1000 and Series 2000 Transmitters ...
Page 106: ...96 Micro Motion Series 1000 and Series 2000 Transmitters ...
Page 172: ...162 Micro Motion Series 1000 and Series 2000 Transmitters ...
Page 178: ...168 Micro Motion Series 1000 and Series 2000 Transmitters ...
Page 190: ...180 Micro Motion Series 1000 and Series 2000 Transmitters ...
Page 268: ...258 Micro Motion Series 1000 and Series 2000 Transmitters ...
Page 280: ...270 Micro Motion Series 1000 and Series 2000 Transmitters ...
Page 281: ......