Instruction Manual
D200124X012
2500 and 2503 Controllers/Transmitters
June 2017
24
D
For Direct Acting 2500, 2500T,
0.2 bar
(3 psig) for a 0.2 to 1.0 bar (3 to 15 psig) output or 0.4 bar (6 psig) for a 0.4 to
2.0 bar (6 to 30 psig) output.
D
For Reverse Acting 2500, 2500T
, 1.0 bar (15 psig) for a 0.2 to 1.0 bar (3 to 15 psig) output or 2.0 bar (30 psig) for a
0.4 to 2.0 bar (6 to 30 psig) output.
6. Visually examine the nozzle and flapper to ensure the nozzle is as square as possible with the flapper. The nozzle can
be realigned by loosening the Bourdon tube mounting screws (key 45, figure 16) and rotating the Bourdon tube
slightly. If the nozzle is realigned, tighten the mounting screws and repeat step 5.
7. Set the process variable to the maximum value of the input range as shown in table 6.
Table 6. Minimum and Maximum Limits for Setting Process Variables
Application
Minimum Limit
Maximum Limit
Liquid Level
Displacer must be completely out of liquid
Displacer must be completely submerged in liquid
Interface
Displacer must be completely submerged in lighter of two
process liquids
Displacer must be completely submerged in
heavier of two process liquids
Density
Displacer must be completely submerged in liquid having
specific gravity of lowest range point
Displacer must be completely submerged in liquid having
specific gravity of highest range point
8. The output pressure should be:
D
For Direct Acting 2500, 2500T
, 1.0 bar (15 psig) for a 0.2 to 1.0 bar (3 to 15 psig) output or 2.0 bar (30 psig) for a 0.4
to 2.0 bar (6 to 30 psig) output.
D
For Reverse Acting 2500, 2500T
, 0.2 bar (3 psig) for a 0.2 to 1.0 bar (3 to 15 psig) output or 0.4 bar (6 psig) for a 0.4
to 2.0 bar (6 to 30 psig) output.
10. If the unit does not contain an indicator assembly, go to the Startup section. If the unit contains an indicator
assembly, change the pointer span by loosening the indicator plate screws (key 41, figure 16, detail of indicator
assembly), slide the plate until the HIGH mark is under the pointer. Tighten the plate screws and go to the Startup
section.
Note
strip base (key 27) right or left along the elongated slotted hole as follows:
D
To
increase
the output pressure span, slide the flexure strip base away from the torque tube shaft.
D
To
decrease
the output pressure span, slide the flexure strip base toward the torque tube shaft.
Retighten the screws. If the flexure strip base has been moved as far as possible and the output pressure span is still
too large or too small, proceed to step 13.
12. Repeat the procedure from step 4 until the required calibration points are obtained.
Note
Any change of the PROPORTIONAL BAND or SPECIFIC GRAVITY adjustment in the following step changes the zero as well as the
output pressure span.