Calibrating the meter (CAL)
Figure 6-9:
The multi-point linearization (noPtS) can be used to linearize the display for non-linear
signals. For instructions on how to use this feature, see
Multi-Point linearization (noPtS)
.
For instructions on how to program numeric values see
1.
Press the Up arrow button to scroll to the Calibration menu (CAL) and press Enter.
2.
The meter displays noPtS. For a linear signal, press the Up arrow. For a non-linear
signal, refer to
3.
The meter displays CAL UNITS. Press Enter to select the input units/pulse. The menu
reads Fun iT.