Reference Manual
00809-0100-4308, Rev BA
Section 4: Configuration
December 2014
Configuration
4.7
Verify Level
Run the Verify Level tool to match the product level reported by the device to a reference
measurement (measured by using for example handgauging). If any difference, the Calibration
Offset parameter will be adjusted as shown in
A minor adjustment using Calibration Offset is normal. There may, for example be a deviation
between the actual tank height and the configured value.
Non-metallic (e.g. plastic) vessels and installation geometry may introduce an offset for the
Upper Reference Point. This offset may be up to ± 2 in. (50 mm). The offset can be compensated
for using Calibration Offset.
Note
Before running Verify Level, make sure that; the product surface is calm, the tank is not being
filled or emptied, and the actual level is well above the probe end.
To run Verify Level, do the following:
1.
From the
Home
screen, select
Configure
.
2.
Go to
Guided Setup > Initial Setup
.
3.
Click
Verify Level
to check your level measurement, and follow the instructions.
Figure 4-15. Calibration Offset
Reported level
Actual level
Positive Calibration
Offset value
Negative Calibration
Offset value
Summary of Contents for 3308A Rosemount 3308 Series
Page 2: ......
Page 6: ...vi Reference Manual 00809 0100 4308 Rev BA Title Page December 2014 ...
Page 12: ...vi Reference Manual 00809 0100 4308 Rev BA Table of Contents December 2014 Table of Contents ...
Page 14: ...2 Reference Manual 00809 0100 4308 Rev BA Section 1 Introduction December 2014 Introduction ...
Page 52: ...40 Reference Manual 00809 0100 4308 Rev BA Section 3 Installation December 2014 Installation ...
Page 80: ...68 Reference Manual 00809 0100 4308 Rev BA Section 5 Operation December 2014 Operation ...
Page 196: ...4 Reference Manual 00809 0100 4308 Rev BA Index December 2014 Index ...
Page 197: ......