MDPS potentiostat probe series module
Located under mainboard cover there are two connectors that can be used to install probe modules. Modules come
pre-installed upon request and may appear different as shown (different configurations). Identify installed modules
to correctly connect probes.
To ensure reliable results probe’s head must be cleaned
on regular basis. This menu enable the embedded
wiper to clean probe’s head.
Cycle Time:
time until next cleaning procedure (from
6hours up to 10 days)
Clean Time:
overall time required to perform cleaning
procedure (from 0 to 999 secdonds)
Restore Time:
time that probe needs to return reliable after
cleaning procedure (from 0 to 999 minutes)
Self Clean Menu
Within calibration menu enable or disable pH &
temperarure compensation. For proper probe’s
calibration please refer to Cl channel
pH & Temperature compensation
Block n. 1
Probe Wire Green
Block n. 2
Probe Wire White
Block n. 3
Probe Wire Brown
Block n. 4
Probe Wire Blue + Wire Yellow
Block n. 5
Probe Black on Block n.8 mainboard (N)
Block n. 6
Probe Red on Block n.9 mainboard (N.O.)
1 2 3 4 5 6