Cl Calibration.
Cl calibration procedure involves probe’s selection, Zero and 2nd Point calibration.
This procedure assumes that instrument is correctly configured and a working Chlorine probe connected and installed
on system. Measurement must be performed using plant water.
Otherwise unattended results may occurr.
Select Probe.
Once into “Calibration Cl” menu move on “Select Probe” then press wheel to enter into probe list. According to system
choose most suitable probe by moving the wheel. Then press to confirm. Note: SCLxx probes series don’t require
ZERO calibration and select probe.
fig. 14
Probe
Scale (mg/l)
ECL 1/2
2.000mg/l Cl2
ECL 1/5
5.00mg/l Cl2
ECL 1/20
20.00mg/l Cl2
ECL 1/200
200.0mg/l Cl2
ECL 2/2
2.000mg/l Cl2O2
ECL 2/20
20.00mg/l Cl2O2
ECL 3/2
2.000mg/l Cl2
ECL 3/10
10.00mg/l Cl2
ECL4,5,6,7
10.00mg/l Cl2 or Br
ECL 8/2
2.000 mg/l Cl2Tot
ECL 8/20
20.00 mg/l Cl2Tot
ECL 9/200
200.0mg/l H2O2
ECL9/2000
2000mg/l H2O2
ECL10/1
0.5mg/l O3
ECL 10/10
10.00mg/l O3
ECL11/200
200.0mg/l peracetic acid
ECL11/2000
2000mg/l peracetic acid
ECL 13
60.00mg/l O2
ECL 17/10
10.00mg/l Cl2O2
ECL 18/10
10.00 mg/l Cl2