Aa and Ab
Using Aa and / or Ab is possible to set a visual alarm with delay for values ≥ or ≤ than set value. Once that Aa or Ab
function is enabled (move wheel over “disable” , press, rotate to “enable”, press again to exit) then main parameters are
ALARM and DELAY.
ALARM.
Alarm condition can be set “ON” or “OFF” and for pH values ≥ or ≤ than set value. Reaching that value an alarm
message will pop-up.
DELAY.
A delay from 0 to 99 minutes can be set before the instrument generates the alarm.
MSG.
when flagged an SMS or EMAIL alarm message will be sent to destination edited in GSM menu.
AD
Using AD is possible to set a maximum dosing time. This alarm prevents connected pump to dose if set time is
reached. Time can be set between 1minute and 9hours and 59minutes. Related setpoint output will be disabled when
AD alarm condition is reached.
MSG.
when flagged an SMS or EMAIL alarm message will be sent to destination edited in GSM menu.